Sunday, October 29, 2023

IEC 62304 Software Maintenance Process

Hi friends!!!


In our previous blogs, we have discussed about the following topics.


IEC 62304 Software System Testing


IEC 62304 Software Unit Implementation & Integration


IEC 62304 Software Architectural & Detailed Design


IEC 62304 Software Requirements Analysis


IEC 62304 Software Development Planning


IEC  62304 Overview & Definitions


IEC 60601-1 Definitions


IEC 60601-1 Classification of Medical Electrical Equipment


IEC 60601-1 General Requirements for Testing Medical Electrical Equipment


IEC 60601-1 Medical Electrical Systems


IEC 60601-1 Protection against unwanted and excessive radiation hazards


IEC 60601-1 General Requirements


IEC 60601-1 Programmable Electrical Medical Systems 


IEC 60601-1 Hazardous situations and fault condition


IEC 60601-1 Construction of Medical Electrical Equipment


IEC 60601-1 Leakage Currents and Patient Auxiliary Currents


IEC 60601-1 Creepage Distance and Air Clearance


IEC 60601-1 Components and Wiring


IEC 60601-1 Mechanical Hazards Moving Parts


IEC 60601-1 Excessive Temperatures


IEC 60601-1 Markings on the Outside of Medical Electrical Equipment


IEC 60601-1 Markings on the Inside of Medical Electrical Equipment


IEC 60601-1 Instability Hazards


IEC 60601-1 Separation of Parts


EU MDR Assessment Routes


EU MDR Timeline, Annex, Chapter


EU MDR Post Market Surveillance


EU MDR Clinical Evaluation


EU MDR Technical Documentation - Part 1


EU MDR Technical Documentation - Part 2


EU MDR General Safety and Performance Requirement


EU MDR Information on the Label


EU MDR Post Market Clinical Follow Up


EU MDR Unique Device Identification - Part 1 


EU MDR Unique Device Identification - Part 2


EU MDR Classification Rules


EU MDR EUDAMED


EU MDR Economic Operators


EU MDR Chapter & Articles - Part 1


EU MDR Chapter & Articles - Part 2


In today's blog, we are going to discuss about IEC 62304.


IEC 62304 states the requirements for Medical Device Software. Satisfying the requirements of IEC 62304 is an important task in order to ensure the safety of patients and efficacy of software used in treating the patients.


Therefore, it is important for Medical Device QA/RA Professionals to understand about Software Maintenance Process requirements as per Clause 6 of IEC 62304. 

 

Software Maintenance Process

The Software Maintenance is performed once the medical device software is released in the market. There are multiple reasons for performing maintenance activities which includes correcting issues & bugs and to keep up with the customer needs etc. IEC 62304 states the requirements for Medical Device Software Maintenance Process.

Software Maintenance Plan

The Software Maintenance Plan shall be established by the manufacturer of the medical device software. The Software Maintenance Plan shall address the following:

The Procedures for addressing the feedbacks once the medical device software is released into the market.

The Criteria for determining whether the feedback can be considered as a problem.

The Problem resolution process & Configuration Management Process.

The Procedures for evaluating and implementing the upgrades, bug fixes, patches, obsolescence of SOUP.

These requirements are applicable for Class A, B & C, medical device software.

 

Problem and Modification Analysis

 

Document and Evaluate Feedback:

The feedbacks of the market released medical device software shall be monitored, documented and evaluated by the manufacturer of the medical device software to determine the existence of problems in the released medical device software.

The Problems identified shall be recorded in Problem Report as per Problem Resolution Process specified in IEC 62304.

The Problem Report shall include deviations from specification and actual or potential adverse events.

The effects of the problem on the safety of the medical device software released into the market shall be evaluated in the Problem Report. The changes needed in the medical device software for addressing the problem shall also be evaluated in the problem report.

The requirement is applicable for Class A, B & C, medical device software.

 

Software Problem Resolution Process & Change Requests:

The Problem Resolution Process as per IEC 62304 shall be used by the manufacturer of the medical device software to address problem reports.

In addition to the Problem Resolution Process, each change requests raised to address the problem shall be analyzed by the manufacturer with respect to its effect on released medical device software and the systems with which it interfaces, and the organization.

Each change requests which modify the released medical device software shall be evaluated and approved by the manufacturer. The Change request means a documented specification about the change.

The Users and Regulators shall be communicated by the manufacturer about any problems in the released medical device software. The Users and Regulators shall also be communicated about the consequences of continued unchanged use of the software.

The requirement is applicable for Class A, B & C, medical device software.

 

Implementation of Modification

Software Development Process as per IEC 62304 need to be identified and repeated by the manufacturer of medical device software for implementing the required software modification.

Modifications made to the software shall be released by the manufacturer of medical device software. The release can be a full re-release of software system or only as a software item comprising of changes and tools to install the changes to an existing system.

The requirement is applicable for Class A, B & C, medical device software.


Now I hope you have acquired some basic knowledge about Software System Maintenance Process as per Clause 6 of IEC 62304. We will meet again with another valuable topic related to Biomedical Engineering, Medical Device, Healthcare.


NOTE: Dear friends!!!... Please do comment a topic related to Biomedical, so that we can discuss it in future blogs.


Check out for this blog about Computed Tomography,  X ray, MRI scan, ECG.


Generations of CT Scanners


COMPUTED TOMOGRAPHY|TECHNOLOGY|ARTIFACTS


x ray machine working principle


components of x ray tube


MRI scan 


MRI With Contrast


Brain MRI


Types of MRI scanners


ELECTROCARDIOGRAM


12 Lead ECG Placement


HOW TO READ AN ECG


Holter Monitor


DON’T FORGET TO FOLLOW THIS BLOG PAGE


DROP YOUR MAIL ID TO GET IMMEDIATE UPDATES


If you find this blog post knowledgeable, 

comment & share it with a friend!